Fluke 718-30US Pressure Calibrator with Pressure Vacuum Pump, -12 to 30 PSI
Pressure Calibrators
Take pressure source and milliamp measurements to calibrate and maintain almost any pressure device.
| Stock No. | FLK665075 |
|---|---|
| Manufacturer | Fluke |
| Part No. | 718 30G |
| Supplied/Stocked By | FLUKE |
Fluke 718-30US Offfers
The Fluke 718 Pressure Calibrator provides a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.
This meter provides a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.
The 718 presents cost-effective solutions for highly accurate pressure measurements over a wide range of temperatures. Connect it with any of the Fluke pressure modules to address a wide variety of gauge, differential, absolute, vacuum and dual (compound) pressure applications.
The Fluke 718 Pressure Calibrator provides a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.
Features
- Error % calculation for quicker pass/fail decision making in the field
- Best in class mA accuracy of 0.015 % for better measurement confidence and workload coverage
- Min/Max/Hold to capture changing measurements
- Switch test feature for quicker and more accurate pressure switch testing
- New 1 and 300 PSI pressure ranges mean fewer extra tools required
- 1, 30, 100 & 300 PSI ranges now available
Applications
- Pressure Instrument Calibration
- Pressure Instrument Maintenance
- Petro-Chemical
- Oil and Gas
- Pharmaceutical
- Water and Wastewater
- Food and Beverage
- Power Generation
Application Selection Guide
Pressure Calibration Selection GuideWhat's included with the Fluke 718 30G
- Fluke 718-30US Pressure Calibrator with Pressure Vacuum Pump
- Protective Yellow Holster with test lead storage
- Instructions Sheet (14 languages)
- Test Leads and Alligator Clip
- Two 9V alkaline batteries
| Range | -12 to 30 PSI, -850 mbar to 2 bar (internal sensor) Over Pressure 3xFS |
|---|---|
| Resolution | 0.001 psi, 0.1 mbar |
| Accuracy | 0.05% full scale |
| Functions | Zero, Min, Max, Hold, Damp, Note: for use with non-corrosive gasses |
| Loop Power Range | 24 V dc |
| Loop Power Accuracy | +/- 10%, Note: Drive; 20 mA into 1000 Ω for battery >6.8V; 700 Ω for battery 5.8 to 6.8 V |
| Measure Accuracy Range | 0 to 24 mA |
| Measure Accuracy Resolution | 0.001 mA |
| Measure Accuracy | 0.025% + 1 count |
| Display | LCD, 5 digit pressure and current simultaneous |
| Operating temperature | -10 to 55°C |
| Operating Altitude | 3,000m max |
| Safety | CSA C22.2 No. 1010.1:1992 |
| Maximum voltage | 30V |
| Size | 216 mm L x 94 mm W x 67 mm (8.5 in. L X 3.72 in. W X 2.61in.) |
| Weight | Approx. 992g (35 oz) |
| Shock & Vibration | Random, 2G, 5-500 Hz, 1m drop test |
| Battery type | Two 9V batteries, ANSI/NEDA 1604A or IEC 6LR61, Battery life: 4 to 20 hours, typical, depending on functions |